Hot spot failure and cause analysis of high and low temperature test chamber

Dongguan manufacturers analyze the hot spots and causes of high and low temperature test chamber operation.

Hot issue 1. Temperature drop is very slow when used

Analysis of the cause: In the above situation, first check whether the studio is dried before the low temperature test, and then do the experiment, or the sample is placed too much, so that the wind in the studio cannot be fully circulated. In addition, after the above reasons are excluded, the inspection is performed. Under the failure of the refrigeration system, this requires the manufacturer to go to the company for professional maintenance.

Hot question 2. Why does the temperature rise after it rises to a certain value?

Analysis of the reasons: mostly because the working environment is not good, the ambient temperature of the equipment room, the distance between the cabinet and the wall of the high and low temperature test chamber is clearly defined in the operating instructions of the test chamber.

Hot issue 3. Do a damp heat test, why the humidity reaches 100%

Analysis of the cause: It may be caused by the drying of the gauze on the wet bulb sensor. The gauze is used for too long. The purity of the water supply will cause the gauze to become hard and not absorb moisture. In addition, check whether the water tank is short of water.

Hot issues 4. Selection criteria for test chambers

Answer: including size (working size, external size), cabinet structure (internal, external material, appearance treatment, internal sealing, moisturizing, etc.), temperature and humidity requirements (temperature, humidity uniformity, fluctuation, temperature rise and fall rate ), the control instrument used in the control system, electrical components, safety protection devices, refrigeration systems, heating systems and air supply systems.

Hot spot problem 5. Temperature control process pressure abnormality

Analysis of the cause: Due to poor heat dissipation, the high pressure side of the compressor will be high. Therefore, check whether the equipment is 30 cm or more away from the wall. If it is not needed, adjust it in time. In addition, the surrounding sealability is checked. If the test chamber is surrounded by a confined space, the ambient temperature will increase and the pressure on the high pressure side of the compressor will be high.

Braille House Number

Some key features of ADA room number signs include:
1. Raised Characters: The room number must be raised or embossed to allow tactile reading by individuals with visual impairments.
2. Braille: ADA signs must include Grade 2 Braille, which is a system of raised dots that allows individuals with visual impairments to read through touch.
3. High Contrast: The color contrast between the background and the characters must be high to ensure readability for people with low vision.
4. Pictograms: In addition to the room number, ADA signs may include pictograms or symbols to provide additional information, such as indicating accessible restrooms or wheelchair accessibility.
5. Mounting Height: ADA guidelines specify the appropriate mounting height for room number signs to ensure they are easily visible and accessible for individuals using wheelchairs or mobility aids.
It is important for public buildings, such as hotels, hospitals, and offices, to have ADA room number signs to ensure equal access and navigation for everyone, regardless of their abilities.

Braille House Number,Braille Sign Plate,Acrylic Plate Braille Signs,Braille Script Numbers

Shanghai Caide Printing Co.,Ltd. , https://www.caidelabels.com